Set the STEP Count – Scroll usin g the encoder  for the total number of times the steps will be processed via the
2.
Menu → Load Control → Step Param. → Count
Set first step value - Press the CC hot key, press the STEP key to enter step edit mode, scroll  to select step 1.
3.
Press the STEP key to switch between step value and time settings. Select step value and tap the encoder or press
enter. The step current value on the bottom right hand corner of the LC D screen will flash, now enter the new amps
value (5.0), and press enter to confirm. Press the STEP key aga in to scroll to the time edit selection and press enter
or tap the encoder to confirm. The step time value on the bottom left hand corner of the LCD screen will flash, now
enter the new time value (800), and press enter to confirm.
Set second step value - Press the CC hot key, press the STEP key to enter step edit mode, scroll  to select step 2.
4.
Press the STEP key to switch between step value and time settings. Select step value and tap the encoder or press
enter. The step current value on the bottom right hand corner of the LC D screen will flash, now enter the new amps
value (10.0), and press enter to confirm. Press the STEP key again to scroll to the t ime edit select ion and press enter
or tap the encoder to confirm. The step time value on the bottom left hand corner of the LCD screen will flash, now
enter the new time value (400), and press enter to confirm.
Set third step value - Press the CC hot key, press the STEP key to enter step edit mode, scroll  to select step 3.
5.
Press the STEP key to switch between step value and time settings. Select step value and tap the encoder or press
enter. The step current value on the bottom right hand corner of the LC D screen will flash, now enter the new amps
value (50.0), and press enter to confirm. Press the STEP key again to scroll to the t ime edit select ion and press enter
or tap the encoder to confirm. The step time value on the bottom left hand corner of the LCD screen will flash, now
enter the new time value (200), and press enter to confirm.
6.
To start the selected STEP mode, press the STEP hot key to initiate the step mode.
7.
To enable the STEP mode, press the INPUT hot key.
8.
The INPUT will be turned off automatically when the program completes the number of step and loops.
NOTE: To edit any entry press the STEP menu key and turn the encoder to increment to the desired step. Use the STEP
to scroll to select the step value or time setting, and press enter or tap the encoder to select. Use the encoder or
numerical keys to change to the desired setting. To program other steps, turn the encoder to increment to the desired
step. Press the STEP hot key and turn input on to reactivate STEP mode.