Osc Range Selection - Ametek Amrel PLA Series Operation Manual

Programmable dc electronic load
Table of Contents

Advertisement

ONCE Causes the STEP to advance only one point after each trigger. Triggers that arrive during a
dwell delay are ignored
AUTO Causes the entire STEP to be executed sequentially after the starting trigger, paced by its
dwell delays. As each dwell delay elapses, the next point is immediately executed.
Command Syntax
Parameters
Query Syntax
Returned Parameters
NOTE: Please refer to the eLoad Programming manual and Operation Manual for further Step Operat ion and
comprehensive Step SCPI Commands.

6. OSC Range Selection

The OSC Range selection now includes DISABLE + DEFAULT, DISABLE + OSC1, DISABLE + OSC2 &
DISABLE + OSC3. With the new selections, the eLoad provides more bandwidth selections for applications
with dc load + ac perturbation. The DISABLE OSC range will stop detection of oscillating conditions; in the
case ac + dc inputs are required for testing applications.
CAUTION
Product used with oscillation protection turned off may cause product failure. Failure occurring as a result of
oscillation application or turn-off of this feature will not be covered under warranty.
Local Control
MENU
LOAD CONTROL➔ OSCILLATE ➔ OSC CONTROL
M470039-01 REV-F
STEP:CURRent:STATe <NR1>
0 | OFF(0) | 1 | ON(1) | 2 | AUTO(2) | 3 | ONCE(3)
STEP:CURRent:STATe?
<NR1>
0=OFF
1=ON
2=AUTO
3=ONCE
➔ DEFAULT
➔ OSC1
➔ OSC2
➔ OSC3
➔ DEFAULT + DISABLE
➔ OSC1 + DISABLE
➔ OSC2 + DISABLE
➔ OSC3 + DISABLE
Page 119