Ametek Amrel PLA Series Operation Manual page 112

Programmable dc electronic load
Table of Contents

Advertisement

CV READBACK LEVEL CALIBRATION
CV readback main level calibration requires a power supply that co vers full range of input voltage of load and is capable
of operating in CV and CC mode. Estimate the power consumption of load, because power supply will go to CC mode
while calibrating the load. Adjust the CC current level to safe limit of load.
Use "" and "" until you see RB CV0 parameter. Then press enter:
CV0
LO=
CNT:
CV0
HI =
CNT:
Press "" to select RB CV1. Then press enter.
CV1
LO=
CNT:
CV1
HI =
CNT:
Press "" to select RB CV2. Then press enter.
CV2
LO=
CNT:
CV2
HI =
CNT:
CC LOW RANGE READBACK LEVEL CALIBRATION
CC readback low range calibration requires a power supply that covers full range of high range input current of load.
Estimate the power consumption of the electronic load, and adjust the voltage input to a safe limit for the load.
Use "" and "" until you see RB CRO parameter. Press enter:
CC0
LO=
CNT:
CC0
HI =
CNT:
Press "" to select RB CC1. Then press enter.
CC1
LO=
CNT:
CC1
HI =
CNT:
Press "" to select RB CC2. Then press enter.
CC2
LO=
CNT:
CC2
HI =
CNT:
This completes the calibration for all the input parameters for the electronic load.
Press "3" and "ENTER" key simultaneously to save all new calibration parameters into the load's memory.
M470039-01 REV-F
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS"
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Page 111