Ametek Amrel PLA Series Operation Manual page 109

Programmable dc electronic load
Table of Contents

Advertisement

Use "" and "" until you see MAIN CR2 parameter. Press enter:
CR1
LO= CNT:
CR1
LO= CNT:
CR1
HI= CNT:
CR1
HI= CNT:
CONSTANT POWER (CP) LOW/MIDDLE/HIGH RANGE, MAIN LEVEL CALIBRATION
CP main level calibration requires a power supply that has enough power output. Fo r example, a PL W6K-60-150 model
needs at least 40V/80A power output to calibrate the unit.
Use "" and "" until you see MAIN CP0 parameter. Then press enter:
CP0
LO= CNT:
CP0
LO=
CNT:
CP0
HI= CNT:
CP0
HI= CNT:
Use "" and "" until you see MAIN CP1 parameter. Press enter:
CP1
LO= CNT:
CP1
LO= CNT:
CP1
HI= CNT:
CP1
HI= CNT:
Use "" and "" until you see MAIN CP2 parameter. Press enter:
CP2
LO= CNT:
CP2
LO= CNT:
CP2
HI= CNT:
CP2
HI= CNT:
CV TRANCIENT LEVEL CALIBRATION
CV transient level calibration requires a power supply that covers full range of input voltage of load and is capable of
operating in CV and CC mode. Estimate the power consumption of eLoad, because power supply will go to CC mode
while calibrating the load. Adjust the CC current level to a safe limit for load.
Use "" and "" until you see TRAN CV0 parameter. Then press enter:
CV0
LO=
CNT:
CV0
HI =
CNT:
Press "" to select TRAN CV1. Then press enter.
CV1
LO=
CNT:
CV1
HI =
CNT:
M470039-01 REV-F
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS"
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS" message