Ametek Amrel PLA Series Operation Manual page 129

Programmable dc electronic load
Table of Contents

Advertisement

9) Saving the Configuration: Menu → PROFILE → RESET→ Enter to update the operating
Min/Max ranges (It needs to reset again once changed back to Master only)
eLoad Configuration – Slave eLoad
1) Manual Range Control:
RANGE → MANUAL
NOTE: It should be high range only in order to have same current setting controlled by IMON from Master unit.
2) Mode Control: CC Mode only
3) EXT.CONTROL: Set EXT control on
4) Set Main Level: Set Current Level to 0Adc
5) Trigger Source: Set Trigger Source External
6) Set Interlock:
a.
Access the Trigger Source Selection Menu from the Utility Menu
LOCAL (MENU)➔LOAD CONTROL➔TRIGGER➔ACTION EXT➔INTLOCK ─
b.
Press Enter to configure TTL low at TRIG_IN to disable Input
c.
Press INPUT Hot Key on which will be flashing and waiting for trigger to turn on later.
7) INPUT always ON
LOCAL (MENU)➔LOAD CONTROL➔BOOT INPUT➔ON
8) Select System Bandwidth:
→ LOAD CONTROL → OSCILLATE → OSC. CONTROL
MENU
→ OSC1 – Rising and Falling Time = 100
The Slew Hotkey must be press again after changing the slew time setting.
The additional filter board at the input terminal must be added in order to keep the stability at CV
mode.
9) Saving the Configuration: Menu → PROFILE → SAVE → 0
Note: Location 0 will be stored in the non-volatile memory of the eLoad and will be loaded each time the
eLoad is rebooted, however it will be combined with the factory default configuration. Other profile
locations 1, 2, & 3 can be used, but will need to be loaded each time the eLoad is rebooted.
M470039-01 REV-F
High only
μs; it can be default 50us if necessary
Page 128