Ametek Amrel PLA Series Operation Manual page 110

Programmable dc electronic load
Table of Contents

Advertisement

Press "" to select TRAN CV2. Then press enter.
CV2
LO=
CNT:
CV2
HI =
CNT:
CC LOW/MIDDLE/HIGH RANGE, TRANSIENT LEVEL CALIBRATION
CC transient level calibration requires a power supply that covers full range of high range input current of load. Estimate
the power consumption of the electronic load, and adjust the voltage input to a safe limit for the load.
Use "" and "" until you see TRAN CC0 parameter. Press enter:
CC0
LO=
CNT:
CC0
HI =
CNT:
Press "" to select MAIN CC1. Then press enter.
CC1
LO=
CNT:
CC1
HI =
CNT:
Press "" to select MAIN CC2. Then press enter.
CC2
LO=
CNT:
CC2
HI =
CNT:
CR LOW RANGE TRANSIENT LEVEL CALIBRATION
CR low range transient level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of electronic load.
Use "" and "" until you see TRAN CR0 parameter. Press enter:
CR0
LO= CNT:
CR0
LO= CNT:
CR0
HI= CNT:
CR0
HI= CNT:
CR MIDDLE RANGE TRANSIENT LEVEL CALIBRATION
CR middle range transient level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.
Use "" and "" until you see TRAN CR1 parameter. Press enter
CR1
LO= CNT:
M470039-01 REV-F
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER
Input measured voltage from Volt.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.