Use "" and "" until you see MAIN CC0 parameter. Press enter:
CC0
LO=
CNT:
CC0
HI =
CNT:
Press "" to select MAIN CC1. Then press enter.
CC1
LO=
CNT:
CC1
HI =
CNT:
Press "" to select MAIN CC2. Then press enter.
CC2
LO=
CNT:
CC2
HI =
CNT:
CONSTANT RESISTANCE (CR0) LOW RANGE MAIN LEVEL CALIBRATION
CR low range main level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of load.
Use "" and "" until you see MAIN CR0 parameter. Press enter:
CR0
LO= CNT:
CR0
LO= CNT:
CR0
HI= CNT:
CR0
HI= CNT:
CONSTANT RESISTANCE (CR1) MIDDLE MAIN LEVEL CALIBRATION
CR middle range main level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.
Use "" and "" until you see MAIN CR1 parameter. Press enter:
CR1
LO= CNT:
CR1
LO= CNT:
CR1
HI= CNT:
CR1
HI= CNT:
CONSTANT RESISTANCE (CR2) HIGH RANGE MAIN LEVEL CALIBRATION
CR high range main level calibration requires a power supply that has high voltage and medium current, but operator has
to estimate the power consumption of load.
M470039-01 REV-F
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Curr.DMM, then press ENTER
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Page 107