CR HIGH RANGE TRANSIENT LEVEL CALIBRATION
CR high range transient level calibration requires a power supply that has high voltage and medium current, but operator
has to estimate the power consumption of load.
Use "" and "" until you see TRAN CR2 parameter. Press enter:
CR2
LO= CNT:
CR2
LO= CNT:
CR2
HI= CNT:
CR2
HI= CNT:
CP RANGE TRANSIENT LEVEL CALIBRATION
CP high range transient level calib ration requires a power supply that has enough power output. For example, a PLW6K-
60-150 model needs at least 30V/90A power output to calibrate the unit. Only qualified personnel should attempt this
calibration, due to the high level of power.
Use "" and "" until you see TRAN CP0 parameter. Then press enter:
CP0
LO= CNT:
CP0
LO=
CNT:
CP0
HI= CNT:
CP0
HI= CNT:
Use "" and "" until you see TRAN CP1 parameter. Press enter:
CP1
LO= CNT:
CP1
LO= CNT:
CP1
HI= CNT:
CP1
HI= CNT:
Use "" and "" until you see TRAN CP2 pa rameter. Press enter:
CP2
LO= CNT:
CP2
LO= CNT:
CP2
HI= CNT:
CP2
HI= CNT:
M470039-01 REV-F
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the front panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM, then press ENTER.
Input measured voltage from Volt.DMM, then press ENTER.
Input measured voltage from Curr.DMM then press ENTER key. A "SUCCESS" message
should be display on the f ront panel LCD.