Ametek Amrel PLA Series Operation Manual page 119

Programmable dc electronic load
Table of Contents

Advertisement

Step Remote Operation
 Step Programming Functions
STEP
This subsystem programs the stepping functions for stepping mode. The number of loops, step values, and dwell t imes
are programmed for a specified operating mode.
[SOURce:]STEP:COUNt
This command sets the number of times that the STEP is executed before it is completed. The command accepts
parameters in the range 1 through 65535, or infinity (0). Program "0 (infinity)" as the command parameter will execute a
STEP indefinitely.
Command Syntax
Parameters
Query Syntax
Returned Parameters
[SOURce:]STEP:CURRent[:LEVel]
This command specifies the current setting for each step.
Command Syntax
Parameters
Query Syntax
Parameters
Returned Parameters
[SOURce:]STEP:CURRent:TIMe?
This command specifies the dwell time for each STEP. Each value of point represents the time in milli-seconds that the
input will remain at the particular STEP point before completing the step. At the end of the dwell t ime, the input of the
electronic load depends upon the following conditions:
■ If STEP:CURRent:STATe ON has been programmed, the input automatically changes to the next
point in the STEP.
■ If STEP:CURRent:STATe AUTO has been programmed, the input waits for a trigger command to
change to the next point in the STEP.
■ If STEP:CURRent:STATe ONCE has been programmed, the input remains at the present level until
a trigger sequences the next point in the STEP.
Command Syntax
Parameters
Unit
Query Syntax
Parameters
Returned Parameters
[SOURce:]STEP:CURRent:STATe
This command specifies how to process the STEP sequencing. The following parameters may be specified.
ON Causes the entire STEP to be executed immediately, paced by its dwell delays. As each dwell
delay elapses, the next point is immediately executed.
M470039-01 REV-F
STEP:COUNt <NR1> | INFinity
0 | INF(0) | 1 TO 65535 | MIN | MAX
STEP:COUNt?
<NR1>, Loop Count
STEP:CURRent <NR1>,<NRf+>
<NR1> point index, 1 through 256
<NRf+> current level for specified point. Refer to model specifications for
range | MIN | MAX
STEP:CURRent? <NR1>
<NR1> point index, 1 through 256
<NRf+> current level for specified point
STEP:CURRent:TIMe <NR1>,<NR1>
st
1
<NR1>, point index
nd
2
<NR1>, 0 to 65535 | MIN | MAX
ms
STEP:CURRent:TIMe? <NR1>
<NR1> point index
<NR1> dwelling time value in ms
Page 118

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents