Ebike Ii Basic / Comfort; Preventive Maintenance; Final Checkout Procedure / Functional Test - GE eBike II Series Service Manual

Table of Contents

Advertisement

B
II
/
e
Ike
BasIC
CoMfort
p
M
reventIve
aIntenanCe
eBike II basic / comfort is a protection class II device (no connection to protective ground)
preventive maintenance has to be performed every 2 years
f
Inal CHeCkout proCedure
Check the setup for stability, adjust with levelling device if required.
Expected result:
Visually inspect the device:
- inspect coat of lacquer
- inspect cables / power cord
- Check bellows on steering column and saddle
Expected result:
Check cranks, pedals and Velcro strips
Expected result:
eBike basic
- Check handlebar adjustment (angle)
- Check saddle height adjustment
- Check that the clamping levers (saddle (height) and handlebar (angle) adjustment) are tight.
Grease the threads of the clamping levers (every 2 to 6 months, depending on applicationpurpose).
(Use of an universal high-performance lubricant such as OKS 470 is recommended)
eBike comfort
- Check electrical saddle adjustment: adjust minimum and maximum saddle height;
then return saddle to mid-level
- Check correct indication of saddle height display
- Check handlebar adjustment (angle)
- Check that the clamping lever for handlebar adjustment (angle) is tight
Expected result:
Using the power cord, connect the device to the power line and turn it on with the power switch
Wait for selftest to end:
- check LCD (e.g. for error messages)
- check in the software if the firmware version is OK
Expected result: Device powers on without errors and FW is OK
- 78 -
/ f
unCtIonal test
No instability
No damage or wear detected
Cranks, pedals and velcro strips OK
Adjustments work and levers are tight, all threads properly greased
eBike II, eBike II L, eBike II EL
Hint
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
2018112-069 Rev. F