The service menu with the following parameters appears:
Baudrate 8N2
Static Load Calibration
Analog Comparison
ECG-Startpulse
NIBP Parameter
Software Update
Dynamic Load Calibration
PC-Mode
Saddle Calibration
NIBP Calibration
NIBP Test
Ergometer Type
System Config.
Error log
Select the desired parameter by use of the right and left
softkeys (↑ ↓) and confirm with SET.
d
efault settIngs
s
ettIngs Menu
default mode
protocols
contrast
load change
language
beep
date/time
ECG type
rpm
pulse display
s
ervICe Menu
Baudrate 8N2
ECG-Startpulse
NIBP Parameter
PC-Mode-Automatic:
Ergometer Type
Analog comparison
System Config.
Saddle engine
NIBP module
Analog interface
60 Hz Power
2018112-069 Rev. F
(with option COM only)
(with option NIBP only)
(eBike II comfort only)
(with option NIBP only)
(with option NIBP only)
(without option NIBP only)
menu
see operator manual
at optimum
+/- 5W
US devices
english
other devices
german
ON
current values
digital
min: 54 / max: 65
OFF
4800
leadtime:
60 sec
pulswidth:
100 msec
leadtime:
60 sec
Resting NIBP:
ON
ON
ER900
20W = 0,2 V / 300W = 3.0 V
eBike basic/L/EL
0
eBike comfort
1
devices w/o NIBP 0
devices with NIBP I
0
US devices
1
other devices
0
eBike II, eBike II L, eBike II EL

Service Menu

Baudrate 8N2
Static Load Calib.
Analog Comparison
ECG-Startpulse
NIBP Parameter
SET
S
election of paraMeterS in Service Menu
Hint
The service menu in the ergometers is available in english
only.
Depending on the actual ergometer configuration, not all
settings are visible.
Leaving the service menu is only possible by switching the
device off.
The sequence of the following descriptions does not
follow the sequence in the service menu.