Error Code Bits Bde (Boodpressure) - GE eBike II Series Service Manual

Table of Contents

Advertisement

e
Bde (B
rror Code BIts
oodpressure
Bit
Error
0
not used
1
cuff loose
2
Micro position
3
Leakage
4
Errror Calibration
5
Error offset
6
not used
7
not used
8
Safety state mode
9
Setup not ready
10
CAN zykl. Timeout
- 136 -
)
Reaction
Display / Info
Warning
Warning
Warning
Module is switched off
Internal error
Module is switched off
Internal error
Module is switched off
Internal error
Module is switched off
Internal error
Module is switched off
Internal error
eBike II, eBike II L, eBike II EL
Troubleshooting
- check correct placement of cuff
at patient (tight)
- check cuff
- check correct placement of
microphone (see User Manual)
- check microphone in cuff
- replace cuff
- check cuff
- check hoses, hose connections
- check compensation container
- replace pneumatic unit
- perform NIBP test
- perform NIBP calibration
- replace BDE PCB
- perform NIBP test
- perform NIBP calibration
- replace BDE PCB
- restart eBike (OFF / ON)
- if error persists:
replace BDE
- restart eBike (OFF / ON)
- if error persists:
check cable ZME <-> BDE
- replace BDE
- restart eBike (OFF / ON)
- if error persists:
check cable ZME <-> BDE
- replace BDE