All the different components are connected with tubes
(one pressure system) and connected to the cuff connector
of the ergometer control terminal (use cuff adaptor).
nIBp t
est
This function is used to check the internal pressure system
(pressure sensor, closeness, leakage etc.).
Select „NIBP Test" in the service menu and confirm with
ENTER (or SELECT).
Using the hand pump, apply a fixed pressure to the system
The actual pressure value is displayed.
Compare the displayed ergometer value with the value of
the manometer.
The maximum allowed deviation is +/- 3mmHg.
- 72 -
compensating container
c
alibration Setup

NIBP Test

NIBP pressure
t
eSt of the internal
If the applied pressure reaches 320 mmHg, the internal
safety valve is opened.
Press „Close valve" (middle soft key) to continue with the
NIBP-Test.
eBike II, eBike II L, eBike II EL
manometer
nibp
123 mmHg
CLOSE VALVE
nibp
preSSure SySteM
Hint
2018112-069 Rev. F
hand pump