For CardioSoft/CS V6.73 and above:
Click onto „COM Port Number" and select an unused
COM port between COM1 and COM32.
Confirm with OK.
Close all open windows.
CoM - I
nterfaCe extensIon
(MaC 5000 st/5500)
The MAC 5000 ST / MAC 5500 use analogue signals
to set the load of the eBike, while the bloodpressure
measurement is controlled with digital signals..
The eBike II is by default only equipped with a digital
interface.
To connect an eBike II with a MAC 5000 ST / 5500 a
special adapter (COM module PN 2008xxx-001) has to be
installed to add an analogue interface to the eBike II.
A jumper adapter has to be placed onto the BKE board to
connect all required signals to the DSUB connector:
Open the housing of the BKE on the bottom side of
the ergometer.
2018112-069 Rev. F
coM
o
pen the houSing of the
eBike II, eBike II L, eBike II EL
(pn 2008
)
Module
xxx
with JuMper and Screw
bKe (
bootoM Side
)
- 33 -