Error Log; Nibp Test / Nibp Calibration - GE eBike II Series Service Manual

Table of Contents

Advertisement

e
rror log
The software of the eBike II ergometers stores internal
error messages in a special memory area (error log file).
These error messages can be checked and cleared in this
menu.
To display the stored error messages select „Show Error
Log" and confirm with ENTER.
The first error message is displayed:
Source:
The unit generating the error
1. Date:
Date of first occurence of this error
1. Time:
Time of first occurence of this error
Err.Code:
Corresponding error code of the software
Quantity:
Number of occurences of this error
By using the arrow keys, the following error messages can
be displayed.
RETURN closes the error log display
The software of the eBike II ergometers stores the signal
strength of each measure and indicates failed measures.
nIBp t
/ nIBp C
est
This menu is visible, if the ergometer is equipped with an
automatic NIBP module and this module is activated.
Test and calibration of the NIBP module is described in the
separate chapter „Calibration / alignment".
2018112-069 Rev. F
alIBratIon
eBike II, eBike II L, eBike II EL

Error log

Err. Log Quantity:
Show Error Logs
Show NIBP Logs
Clear Error Logs
Clear NIBP Logs
ENTER
e
rror log MeMory
Error No.:
1 / 3
Source:
ZME
1. Date:
02.09.2008
1. Time:
14:20
Err.Code:
0008
Quantity:
3
RETURN
d
iSplaying a Single error
NIBP
failed:
Mic.
level 1:
Mic.
level 2:
Mic.
level 3:
Mic.
level 4:
Mic.
level 5:
RETURN
nibp
log
2
N.N.
N.N.
N.N.
N.N.
N.N.
N.N.
- 53 -

Advertisement

Table of Contents
loading

Table of Contents