Alignment „Ouput: Analog Load Value; Alignment „Output: Analog Heartrate; Alignment „Ouput: Analog Systole - GE eBike II Series Service Manual

Table of Contents

Advertisement

a
„o
: a
lIgnMent
uput
The output voltage representing a special load value of the
ergometer is defined with this setting.
It is possible to adjust the voltage and / or the load value.
Select the parameter by use of the arrow keys and confirm
with „Enter"
Adjust the value by using the arrow keys and store the
setting with SET.
Setting range:
GAIN:(voltage)
20 mV to 10000 mV in steps of 2 mV
GAIN (load)
20 to 999 Watt in steps of 10 Watt
a
„o
: a
lIgnMent
utput
The output voltage representing a special heartrate is
defined with this setting.
It is possible to adjust the voltage and / or the heartrate
value.
The setting is performed as described above under
Alignment „Ouput: Analog Load value"
Setting range:
GAIN (voltage):
GAIN (heartrate):
a
„o
: a
lIgnMent
uput
The output voltage representing the systolic value is
defined with this setting.
It is possible to adjust the voltage and / or the systolic
value.
The setting is performed as described above under
Alignment „Ouput: Analog Load value"
Setting range:
GAIN (voltage):
GAIN (systole):
- 56 -
l
"
nalog
oad value
H
"
nalog
eartrate
500 to 2000 mV in
2 mV steps
50 to 250 bpm in
10 bpm steps
s
"
nalog
ystole
500 to 2000 mV in
2 mV steps
50 to 250 mmHg in
5 mmHg steps
eBike II, eBike II L, eBike II EL
OUTPUT 1
Load
GAIN
GAIN
ENTER
OUTPUT 1
Load
GAIN
GAIN
SET
a
lignMent of analog output
load
OUTPUT 2
Heart
GAIN
GAIN
ENTER
OUTPUT 3
Systole
GAIN
GAIN
ENTER
3000 mV
300 Watt
3000 mV
300 Watt
"
1000 mV
100
1000 mV
100 mmHg