GE eBike II Series Service Manual page 202

Table of Contents

Advertisement

C
B
p
leanIng tHe
lood
r
M
eMovIng tHe
ICropHone
Pull the end of the cuff through the metal clasp and fold
out the cuff.
Pull on the short Velcro tab to open the microphone pock-
et and carefully remove the microphone.
C
leanIng
Clean the cuff and tubing with a moist cloth.
You can use a dishwashing liquid or mild soap water
(no cleaning agents containing alcohol).
Clean the microphone with a cloth moistened with alcohol
or soap water.
Allow the microphone to dry before reinserting it in its
pocket.
d
IsInfeCtIon
For disinfection, spray a disinfectant sparingly on the cuff,
the tubing and the microphone.
After the contact time indicated by the manufacturer, wipe
all components dry.
Only the following disinfectants are approved for disinfec-
tion:
Schülke & Mayr GmbH:
Antifect ® AF, FF, FD 10
Terralin ® (0,5 %)
Quartamon Med ®
B. Braun Melsungen AG:
Hexaquart plus ® (0,5 % / 5,0 %)
Hexaquart S ® (1,5 % / 5,0 %)
Meliseptol ®
Melsept SF ® (0,5 % / 5,0 %)
ECOLAB:
Incidin Foam ®
I
M
nsertIng tHe
ICropHone
Slip the microphone into the pocket, the metal side facing
the arm.
Guide the microphone cable out of the pocket and to the
right of the Velcro tab. Then close the tab.
Fold the end of the cuff over and introduce it into the
metal clasp.
- 202 -
C
ressure
uff
eBike II, eBike II L, eBike II EL
r
M
eMoving the
icrophone
Warning
• Equipment Damage •
Cuff, microphone and tubing may not under any circum-
stances:
- be immersed in liquids
- be cleaned in a water bath or in running water.
Hint
Strictly observe the manufacturer's instructions for use.
i
M
nSerting the
icrophone
2018112-069 Rev. F