Electrical Safety - GE eBike II Series Service Manual

Table of Contents

Advertisement

e
s
leCtrICal
afety
Current leakage test results meet requirements?
Perform electrical safety checks when indicated.
All indicated electrical safety checks require a pass/fail indication for steps performed.
Record the measurement values in your debrief.
Step
Earth Leakage Current
1.
2.
Neutral open, Forward Polarity
3.
Neutral open, Reverse Polarity
4.
Enclosure Leakage Current
1.
2.
Neutral open, Forward Polarity
3.
Ground open, Forward Polarity
4.
Ground open, Reverse Polarity
5.
Neutral open, Reverse Polarity
6.
Patient Leakage Current to Ground (3)
1.
2.
Neutral open, Forward Polarity
3.
Ground open, Forward Polarity
4.
Ground open, Reverse Polarity
5.
Neutral open, Reverse Polarity
6.
Ground Continuity
1.
AC mains power cord ground prong to
exposed metal surface (ground lug)
(1) NC =
Normal Condition
SFC =
Single Fault Condition
N/A =
Not Applicable
(2) UUT =
Unit Under Test
(3) Test applies for eBike II EL with bloodpressure unit only - measuring point is the microfone connector
2018112-069 Rev. F
Condition
(1)
Forward Polarity
NC
SFC
SFC
Reverse Polarity
NC
Forward Polarity
NC
SFC
SFC
SFC
SFC
Reverse Polarity
NC
Forward Polarity
NC
SFC
SFC
SFC
SFC
Reverse Polarity
NC
NC
eBike II, eBike II L, eBike II EL
UUT - ON
Result
(2)
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ mΩ
Pass/Fail
Leakage
Current limits
500 μA
1.000 μA
1.000 μA
500 μA
100 μA
500 μA
500 μA
500 μA
500 μA
100 μA
100 μA
500 μA
500 μA
500 μA
500 μA
100 μA
Resistance
Less than
200 mΩ