5.4 Manual Calibration

A/D channel Manual-Calibration
Step 1: Click the Manual A/D Calibration tab to show the A/D channel
manual calibration panel. Before calibrating, acquire the
reference voltage from a precision standard voltage reference.
Go to the Range form, select a channel and the target voltage
range according to the input voltage value from a precision
standard voltage reference (Fig. 5-13).
Note:
The input voltage value you selected from a precision standard voltage
n
reference needs to correspond with the one that the MIC-3716 can read.
The input voltage will be analog so the computer will convert the
n
voltage to a digital voltage; therefore, the input voltage value you
selected from a precise standard voltage reference needs to correspond
with the one that the MIC-3716 can read. For example, if the input
range is 0 ~ 5V, the input voltage should be 2.9992V not 3V.
62