Error Message List

Please refer to the error messages list table as below to
find the corresponding message description:
Error
Error Message
Code
0
No error
–410 Query INTERRUPTED
–350 Query overflow
–321 Out of memory
–224 Illegal parameter
value
–223 Too much data
–222 Data out of range
–220 No matched module
–200 Execution error
–171 Invalid expression
Description
Indicates that a condition causing an
INTERRUPTED query occurred (see IEEE
488.2, 6.3.2.7)
Indicates the SCPI remote interface error
queue overflowed.
Indicates an internal operation needed
more memory than that was available.
Indicates you sent a parameter for this
command that is NOT allowed.
Indicates a legal program data element of
block, expression or string type was
received that contained more data than
the device could handle due to related
device-specific requirements or memory.
Indicates a legal data was parsed but
could not be executed because of the
interpreted value was outside the legal
range defined by the analyzer. The
displayed results may be clipped.
Indicates no matched measurement or
mode found.
This is a generic execution error for
devices that cannot detect more specific
errors. The code indicates on those
execution errors defined in IEEE 488.2,
11.5.1.1.4 has occurred.
Indicates the data element was invalid, for
example, unmatched parentheses, or an
illegal character.
Error Messages
Error Message List