• Limit (*.LIM)
A limit file records the limit settings. You can edit the
limit file and transfer this file between the N9340B
and PC software.
• CSV (*.CSV)
A CSV file records the trace data and available for
review on PC.
• Setup (*.SET)
A setup file records the system setting information,
such as language, date/time, and power saving mode.
This file type is not used for test parameter setup
information.
Save Path
Press [SYS] > {File} > {File Setup} > {Save Path} to select
a path from the local memory or external USB memory
device.
Saving a file
Refer to the following three steps to save a file:
1 Press [SYS] > {File} > {File Setup} to select the save
path and file type.
2 Edit a file name.
A file name can consist of letters and digits. A single
key stroke on the numerical keypad inputs a digital
number; and consecutive key stroke selects and
inputs a letter.
3 Press {Save} to save the file.
When a file saving completes, the message "File Saved"
displays on the bottom line of the screen.
When the saving path to USB, files are automatically saved into a
NO TE
folder named as N9340DATA in the root directory of the USB
memory stick.
If the USB memory stick connection can not be updated, remove
the USB memory stick, then cycle the power and re-insert the
USB memory stick.
System and File
File
33