Getting Started
Making a Basic Measurement
Viewing a Signal on the Analyzer
1 Use a signal generator to generate a CW signal of 1.0
GHz, at a power level of 0.0 dBm.
2 Press [SYS] > {Setting} > {PwrOn/Preset} > {Preset
Type} and select DFT to toggle the preset setting to
factory-defined status.
3 Press the green [Preset] key to restore the analyzer to
its factory-defined setting.
4 Connect the generator's RF OUT connector to
analyzer's RF IN connector.
5 Press [FREQ] > 1 > {GHz} to set the analyzer center
frequency to 1 GHz.
6 Press [SPAN] > 5 > {MHz} to set the analyzer
frequency span to 5 MHz.
7 Press [MARKER] > {Peak Search} to place a marker
(M1) at the highest peak (1 GHz) on the display.
The Marker amplitude and frequency values appear in
the function block and in the upper-right corner of the
screen.
Use the front-panel knob, arrow keys, or the softkeys in
the Peak Search menu to move the marker and read out
the value of both frequency and amplitude displayed on
the screen.
Figure 2-1View a signal (1 GHz, 0 dBm)