Ref Oven (Opt. 1E5) Replacement; Tools Required; Removal Procedure; Replacement Procedure - Keysight Technologies E5061B Service Manual

Network analyzer
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Replacement Procedure

Ref Oven (Opt. 1E5) Replacement

Ref Oven (Opt. 1E5) Replacement

Tools Required

Removal Procedure

Refer to
Step 1. Remove the Outer Cover as described in
page
Step 2. Remove the three 9/16 inch nuts and the three washers (item 1) from the
BNC connectors.
Step 3. Remove the seven TORX T10 screws (item 2) fastening the rear panel.
Step 4. Disconnect the two cable connectors (item 3) from the Ref Oven.
Step 5. Remove the three TORX T10 screws (item 4) fastening the Ref Oven.
Step 6. Lift the Ref Oven.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Set a torque to 30.4 in-lb for box torque wrench when fasten the 9/16 inch
nuts and the washers (item 1).
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws
(item 2, and item 4).
84
torque screwdriver, T15 and T20
torque screwdriver, T10 (set to 9 in-lb)
box torque wrench, 9/16 inch (set to 30.4 in-lb)
Figure 4-16
for this procedure.
55.
"Outer Cover Replacement" on
Keysight E5061B Network Analyzers