Troubleshooting

How to exit from the E5061B Measurement View

How to exit from the E5061B Measurement View
You need to exit from the E5061B Measurement View to perform some
troubleshooting. The following is the procedure to exit from the E5061B
Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5061B
rear panel.
Step 2. Turn the instrument on.
Step 3. Press
Step 4. Click Service Menu - Exit. Exit dialog box is as shown in
Figure 2-1
Exit dialog box
Step 5. Click OK in Exit Menu. Then the E5061B exit the Measurement View,
then.windows desktop screen appears.
If you wish to return to the Measurement View, double-click "Network
Analyzer" icon.
If you need to shut down the E5061B and again turn on, perform in
accordance with the following procedure.
a. To get "Start" menu bar displayed, move the pointer to the bottom of the
b. Click "Start" and "Shut Down..." in the pull down menu. "Shut Down
c. Select "Shut down" button in the pull down menu.
d. Click "OK" button in the dialog box.
20
key.
screen with mouse.
Windows" dialog box opens.
Figure
2-1.