Diagnostic Test Failure Troubleshooting - Keysight Technologies E5061B Service Manual

Network analyzer
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Troubleshooting
Troubleshooting Using Diagnostic Test
Figure 2-16 Setup dialog box

Diagnostic Test Failure Troubleshooting

Table 2-3
the relationships of failed tests to probable faulty board assemblies. If the
instrument fails the diagnostic test, replace the faulty board assembly as
shown in
Table 2-3
Diagnostic tests failure troubleshooting information (Option 3L5)
Test
No.
1
36
The test result file named "resultDT.txt" is created on the
drive D (D:\Keysight\Service\Log\) of the E5061B after "Exit"
button of the Main Menu is pressed. The text file can be read
and edited with a PC.
and
Table 2-4
represents the contents of the diagnostic tests and
Table
2-3.
Test group
DC-BUS Test
Failured
Probable faulty board assembly
Test
A1/A11
A2
Source
Receiver
Module
Module
ALL
###
A1/A11
###
#
Source
Module
A2
###
Receiver
Module
A9 LF
#
Source
Bias
Module
Keysight E5061B Network Analyzers
A9 LF
Source
Bias
Module
##
#
#
###