2.4. Water supply and PIEZON

Without Bottle:
PIEZON
& AIRFLO W
use external water supply.
®
®
1
The CLIP+CLEAN shall be
previously cleaned and disinfected
before use.
Non-disinfected CLIP+CLEAN may
contaminate the device.
The PIEZON
®
bottle is compatible with the following solutions:
Avoid unwanted chemical reactions or ingestion of residues of solution. Always rinse the liquid
circuit with drinking water before using a different solution or undergo a regular treatment requiring
water only.
For treatment application use the percentage of solution as per discretion and per local country
regulations and recommendations.
Solution
Sodium Hypochlorite
Chlorexidine
Carbanilides
Povidone iodine
Quaternary ammonium compound
Cetylpyridin chloride
Citric acid
Physiological water (NaCl %)
Ethanol
Isopropanol
Do not sterilize the PIEZON
Use only ambient temperature active disinfectant and cleaning agents.
ENGLISH
®

bottle

2
Place the CLIP+CLEAN into the
device's bottle receptacle for dust
protection.
Concentration
Up to 5%
Up to 1%
Up to 0.05%
Up to 12%
Up to 0.7%
Up to 1%
Up to 5%
Up to 0.9%
Up to 5%
Up to 5%
bottle and its nozzle cap by steaming or dry thermal reprocessing.
®
FB-618/AU – rev.E – ed.2019/08
With Bottle connected:
AIRFLOW
uses external water supply.
®
PIEZON
®
uses bottle liquid supply.
Connect the PIEZON
Only use the PIEZON
®
(transparent) for disinfectant solutions.
bottle
®
bottle EG-111
14/48