Cleaning & Reprocessing; Water Line Cleaning - EMS AIRFLOW Prophylaxis Master Instructions For Use Manual

Hide thumbs Also See for AIRFLOW Prophylaxis Master:
Table of Contents

Advertisement

5. CLEANING & REPROCESSING

5.1. Water Line Cleaning

Keeping the device's water lines clean is mandatory also to
prevent patient infection.
NIGHT CLEANER cleans the fluid lines of all AIRFLOW
Prophylaxis Master units.
NIGHT CLEANER removes and protects from algae and
limescale, after longer idle times or heating of the process
water.
The water supply hose and related device connection will
not be cleaned by this procedure.
Each morning before the first patient: Rinsing
1
Place a fully filled water
bottle onto the device
To reduce the risk of
ingestion of the cleaning
agent by the patient, always
use a fully filled 800ml water
bottle.
Risk of ingestion of the cleaning agent. Check that no more blue residue of NIGHT CLEANER is
flushing out of the cord. Otherwise, repeat the rinsing procedure.
Always empty out and wash the water bottle used for rinsing before any new use. EMS
recommends a weekly use of a bottle cleaning agent (e.g. BC-San 100 from Alpro Medical GMBH).
Risk of ingestion of residue of cleaning agent. During rinsing, a small quantity of cleaning agent
flows back into the water bottle.
ENGLISH
2
Set water to 10
Turn the device ON
Set both water
regulators to 10 to ensure
optimal rinsing.
FB-618/AU – rev.E – ed.2019/08
®
3
Hold both cords over a
sink with CLIP+CLEAN
Contamination prevention:
Do not make any
contact between the sink
and the cords.
WATER SUPPLY
BY HOSE EG-110
NIGHT
CLEANER
4
Press the pedal once,
release, and then wait 1
minute
The white and blue
countdown indicates
remaining time.
Cleaning can be paused
and reset by pressing the
pedal again.
23/48

Hide quick links:

Advertisement