6. MAINTENANCE & TROUBLESHOOTING
6.1. AIRFLOW
®
In case of a clogged handpiece and before the reprocessing of AIRFLOW
Handpieces.
Rinse through central lumina in
the normal direction of flow (no back
rinsing) using Easy Clean with a
disposable syringe filled with more
than 2 ml of drinking water
The Easy Clean tool can be thermally disinfected and also sterilized at up to 135°C in the autoclave.
6.2. AIRFLOW
®
In case of leakage at the AIRFLOW
connection with the AIRFLOW
of the cord with the spare provided in the EL-651 Kit
located in the
6.3. PIEZON
®
light guide check & replace
The light guide loses its transparence after undergoing repeated reprocessing cycles. Check the
transparency of the light guide every month and do the following:
1.
2.
3.
4.
ENGLISH
Handpiece powder unclogging
Easy Clean
Provided in your
handpiece leakage
®
cord, replace the o-rings
®
application box.
®
Remove the tip and unscrew the handpiece nose cap by hand.
Take off the light guide and inspect it.
Place in a new light guide AB-340 (provided in the
Screw the nose cap on again, by hand only.
FB-618/AU – rev.E – ed.2019/08
Application box
®
Blow air to dry.
handpiece
®
and PERIOFLOW
application).
®
34/48
®