General Precautions - EMS AIRFLOW Prophylaxis Master Instructions For Use Manual

Hide thumbs Also See for AIRFLOW Prophylaxis Master:
Table of Contents

Advertisement

Applied Parts
The following items are Medical Device Applied Parts:
AIRFLOW
®
(EL-308) Handpiece
PERIOFLOW
®
PIEZON
®
(EN-060 and EN-061) Handpieces
Applied Parts, under certain operating conditions, may exceed 41°C of temperature and reach
a maximum temperature of 51°C.

1.7. General Precautions

USE EMS ACCESSORIES ONLY!
DO NOT use this device in the presence of flammable anesthetics or oxidizing gases (such as
nitrous oxide (N2O) and oxygen) or in close proximity to volatile solvents (such as ether or alcohol),
as explosion may occur.
DO NOT store the powder near acids or heat sources.
TAKE the following precautions to prevent any adverse events to the patient and/or to the user
in case of electromagnetic disturbances:
Always
refer
Electromagnetic Compatibility".
In case of a wireless pedal malfunction, presumably caused by electromagnetic
disturbances, use the wired pedal instead.
In case of a device malfunction, presumably caused by electromagnetic disturbances, first
verify the cabling, and then move any portable RF communications equipment and mobile
devices placed nearby as far away as possible to rule out interference.
Stop using the device if the electromagnetic disturbances persist and contact EMS
technical support for assistance.
ENGLISH
(EL-354) Handpiece
PS Instrument
The use of any other accessories could lead to
patient injury, malfunction or damage to the device
to
the
information
FB-618/AU – rev.E – ed.2019/08
listed
in
the
chapter
"
6/48

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AIRFLOW Prophylaxis Master and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents

Save PDF