EMS AIRFLOW Prophylaxis Master Instructions For Use Manual page 18

Hide thumbs Also See for AIRFLOW Prophylaxis Master:
Table of Contents

Advertisement

PIEZON
power setting
®
The unit is equipped with NO PAIN
according to the load applied to the instrument.
The following table shows the maximum output power as per user power setting:
PIEZON
®
Power
Risk of tip breakage: with ENDO files, do not exceed 2.5W (power setting "3" max.)
AIRFLOW
pressure setting
®
Both the PLUS and CLASSIC powder chambers have an integrated dynamic pressure regulator
that automatically set the optimal pressure range for the selected powder chamber and related
powder type as detailed in chapter "Powder Chambers".
The following table shows the static and approximate dynamic pressures
chamber and user power setting:
AIRFLOW
®
Pressure
Dynamic pressures depend on handpiece and powder type too. The listed pressures are for information
3
purpose and referring to the commonly used EL-308 AIRFLOW
powders.
ENGLISH
®
technology which provides an adaptive response in function
Power Setting
0
Max Output Power [W] 0
Pressure Setting
0
Static [Bar]
/
CLASSIC dynamic [Bar]
/
PLUS dynamic [Bar]
/
FB-618/AU – rev.E – ed.2019/08
1
2
3
4
5
0.4
1.2
2.1
3.0
3.9
1
2
3
4
2.5
2.7
3.0
3.2
3.5
1.9
2.1
2.3
2.6
2.8
1.5
1.7
1.9
2.0
2.2
Handpiece with DV-082 and DV-048
®
6
7
8
9
10
4.8
5.6
6.4
7.2
8.0
as per selected powder
3
5
6
7
8
9
3.7
4.0
4.2
4.5
3.0
3.2
3.5
3.7
2.4
2.6
2.7
2.9
10
4.7
3.9
3.1