6.4. Handpiece cord replacement

Disconnect the mains plug for purposes of maintenance and in case of malfunction.
Depressurize the powder chamber before disconnecting the AIRFLOW
In case of persistent malfunction or damage to the PIEZON
the part can be easily replaced by the user. Follow the directions for replacement provided with the
spare part supply.
2
1

6.5. Monthly check

Each month check both air and water filters for cleanliness.
Disconnect the mains plug for purposes of maintenance and in case of malfunction.
No maintenance is allowed while in use with a patient.
ENGLISH
Filter color has to be white without significant visible impurities. If not,
replace the filter.
Check
If the water filter needs to be changed more than 3 times a year,
please check the quality of your water line.
water and air
Air filters usually remain cleaner for longer periods of time. Replace
filter cleanliness.
once a year. (The yearly maintenance service includes the
replacement of both filters.)
Good
1.
2. Disconnect the water hose by pulling it off the connector.
3. Pull the filter off by hand or by using a small flat screwdriver.
Worn-out
4. Replace with a new filter and reconnect the hose.
FB-618/AU – rev.E – ed.2019/08
®
or AIRFLOW
Handpiece cord disconnecting procedure:
1.
Unlock the cord system by pushing the lock
switch to the front (Switch located under the
device).
2.
The cord system is now unlocked and can be
removed by pulling it.
Disconnect the power cord from the grid first.
cord.
®
®
Handpiece cord system,