Preparation
Manual pre-cleaning is required:
Immediately after use, rinse the lumen(s) line of the handpiece/instrument with water for 20
seconds. Coarse soiling must be removed immediately after application.
For AIRFLOW
and PERIOFLOW
®
for both lumens (water and powder) clearage before proceeding.
For PIEZON
:
®
Remove any installed instrument.
Remove the nose cap and separate aside the light guide and the O-ring.
For Endochuck file holder:
Remove any installed file.
Remove the screw and separate aside the small O-ring.
For any Instrument mounted on the CombiTorque:
Separate aside the Instrument and the CombiTorque.
Safely transport to the reprocessing area to avoid any damage to the parts and contamination to
the environment and to the people involved in the reprocessing process.
Cleaning shall need to be performed within 1 hour from the use.
ENGLISH
: always carry out handpiece powder unclogging and check
®
FB-618/AU – rev.E – ed.2019/08