Manual Cleaning with Ultrasound Bath
DO NOT USE WITH HANDPIECE BODY
The following validated process is intended
ONLY for the EMS PIEZON Instruments and file
holders:
Remove away any surface soiling (gross
contamination) on the product with a
wetted cloth and tap water.
For part having lumens, flush all lumens
with a spray gun (water jet gun, with a
static water pressure of 2 bar) with cold
tap water for 15 seconds.
Using
an
appropriate
sonication, immerse the part in 0.5%
solution of neodisher MediClean with
deionized water and sonicate for 10
minutes at 40°C (Bandelin, Sonorex
1028K, 35kHz). Taken care that all
lumens are filled with cleaning solution
(use a syringe if needed). All surfaces
must be moistened.
Rinse all lumens flushing with a spray
gun (water jet gun, with a static water
pressure of 2 bar) with cold deionized
water for 15 seconds, and further rinse
the whole part under cold running
deionized water for 10 seconds.
Using an air pistol (compressed air) fully
dry the lumen and the whole part until no
more residues of water are present
(visible or detectable).
ENGLISH
sieve
for
using the Mobile Injector Unit (Rack) Miele E429
FB-618/AU – rev.E – ed.2019/08
Example of correct placement of the parts
in the WD Miele Professional G 7836 CD
29/48