3. DEVICE USE

3.1. Interfaces

NORMAL
Press gently
ON/OFF-mode
1
Standby
Powder chamber
pressurization /
2
depressurization
Power setting
3
PIEZON
®
water
4
AIRFLOW
®
water
5
Pedal (normal)
6
Pedal BOOST
7
(Only on the wireless pedal)
ENGLISH
BOOST
Press hard
7
6
ON: the device goes into operating mode.
OFF: the device reverts back to standby.
(After 1 hour of inactivity, the unit switches to off-mode standby)
Powder chamber is pressurized or depressurized.
A white light illuminating the powder chamber will turn on when pressurized.
During chamber depressurization, the AIRFLOW
white light will turn off at the end of the process.
Entering Standby mode: The powder chamber depressurizes automatically.
Powder chamber depressurization may take up to 10 seconds to complete.
During this time, it is recommended that you leave the AIRFLOW
holder with the nozzle facing down to avoid spraying the purged air and residual
powder upwards.
Place your finger in the groove below the numbers to adjust AIRFLOW
and PIEZON
power:
®
0 (water only, blue indicator)
10 (Maximum)
Memorization of the preselected settings.
Sets the PIEZON
®
water flow rate.
Sets the AIRFLOW
®
water flow rate.
Press the edge of the pedal for normal operation.
The pedal is deactivated when both handpiece cords are placed in their holders.
Pressing hard on the center of the wireless pedal activates power boost.
For easy boost activation: leave the foot on the pedal and put the heel up.
FB-618/AU – rev.E – ed.2019/08
1
2
4
3
®
cord will automatically purge and the
5
Handpiece in its
®
air pressure
®
17/48