EMS AIRFLOW Prophylaxis Master Instructions For Use Manual page 24

Hide thumbs Also See for AIRFLOW Prophylaxis Master:
Table of Contents

Advertisement

Between each patient
Overall cleaning and disinfection
Clean the external surface of the device
with a cloth and alcohol
Clean the unit only with an alcohol-based (ethanol,
isopropanol), colorless disinfectant.
Never use scouring powder or an abrasive sponge. It
will damage its surface.
End of day: Overnight cleaning
Use only EMS NIGHT CLEANER as a cleaning agent.
Other products may damage or not clean the unit, and cause patient intoxication.
1
Place the NIGHT
CLEANER bottle onto
the device
Before placing, remove
CLIP+CLEAN from the device.
Each cleaning consumes 30ml
of NIGHT CLEANER.
Before cleaning, check
that the liquid level is above the
black flange of the bottle's
neck.
ENGLISH
2
Set water to 10
Turn the device ON
Set both water regulators
to 10 to ensure the flow of the
cleaning agent.
FB-618/AU – rev.E – ed.2019/08
Reprocess handpieces and instruments
See the specific following chapters.
Risk of contamination. Always disinfect the bottom
and top areas of device air connections.
3
Hold both cords over a
sink with CLIP+CLEAN
Contamination prevention:
Do not make any contact
between the sink and the
cords.
CLIP+CLEAN shall be
reprocessed after each use.
4
Press the pedal once,
release, and then wait 1
minute
The white and blue countdown
indicates remaining time.
Cleaning can be paused and
resumed by pressing the pedal
again.
Once completed, leave the
NIGHT CLEANER bottle on the
device overnight at least 12h
minimum.
24/48