Disinfection
Any part can be disinfected manually or automatically by a disinfector.
HIGHLY RECOMMENDED!
Automated Disinfection
EMS highly recommends the use of an ISO
15883 compliant automatic washer-disinfector
(WD) for an optimal effectiveness and part service
life
All EMS parts can undergo an Automated Thermal
Disinfection in a washer-disinfector compliant with
the ISO 15883 standard.
The following validated (For example on Miele
Professional G 7836 CD using the Rack Miele E429)
process can be used to achieve an A
Execute a 3 minutes thermal disinfection (final
rinse) with deionized water at 93°C minimum.
Drain.
Drying for at least 20 minutes at 100°C.
Also special instructions and warnings of the
manufacturer of the washer disinfector have to be
followed.
In case of use of a chemical disinfectant, follow
carefully the instructions provided by the disinfection
solution manufacturer.
Example of correct placement of the parts
in the WD Miele Professional G 7836 CD
using the Mobile Injector Unit (Rack) Miele E429
ENGLISH
ALTERNATIVE
Manual Disinfection
DO NOT use any Ultra Sound Bath disinfecting
procedure with the PIEZON
PERIOFLOW
products.
ASP CIDEX OPA
used not diluted and within its Use-Life and Shelf-Life,
respecting the manufacturer's Warnings and precisely
following its Instruction for Use.
Disinfectant shall be poured into a proper tray having
secure lid. Tray size shall have to allow the complete
submersion of the device and to facilitate the evacuation
of air from the inside.
Level of 3000:
0
Warm up the disinfectant to at least 20°C.
Immerse the product completely, and if applicable, fill all
lumens and eliminate air pockets by means of a syringe
fulfilled of disinfectant (to guaranty full internal lumen
contact with the disinfectant).
Close the tray with the secure lid.
Wait for at least 5 minutes and be sure that the
temperature of the disinfectant solution doesn't drop
below 20°C.
The ASP CIDEX OPA disinfectant requires a total of
three (3) rinses, with large volumes of fresh water to
properly remove its residues. Sterile water shall be
poured into proper rinsing trays. Tray size shall have to
allow the complete submersion of the device and to
facilitate the evacuation of air from the inside.
Finally, fully dry it internally and externally.
Always use fresh volumes of water for each rinse. Do not
reuse the water for rinsing or any other purpose. Residues of
disinfectant may cause serious side effects.
FB-618/AU – rev.E – ed.2019/08
Handpieces: it may destroy the
®
®
0.55% disinfectant solution shall be
For each of the 3 rinses do:
- Flush all lumens with large volumes (not less than
100 ml) of rinse water. Use a syringe or a water jet
pistol to correctly accomplish this task.
- Keep the device totally immersed for a minimum of
1 minute in duration.
- Remove the device from the rinsing tray and
discard the rinse water.
, AIRFLOW
and
®
®
30/48