8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
56
8.1.5.4

Refitting the mirror sleeve

➢ Carefully refit the mirror sleeve until it locks in place.
NOTE
There is a risk of damaging the camera window or the coated
sapphire glass if the mirror sleeve is not pushed straight toward the
front.
➢ The mirror sleeve must not come into contact with the camera
window.
➢ Push the mirror sleeve straight toward the camera body; do not tilt
it.
➢ Carefully refit the mirror sleeve until it locks in place.
8.1.5.5

Pre-cleaning of the mirror sleeve

Clean the CEREC Omnicam immediately after use as follows:
1. Carefully wipe the mirror sleeve while it is on the camera so that no
dirt whatsoever can remain stuck and can harden on the surface of
the mirror sleeve. Use an absorbent cotton-based gauze for this
with water and neutral soap or a lint-free cloth that has been soaked
in one of the cleaning agents listed under Cleaning agents and
disinfectants  [→ 52].
2. Next dry the mirror sleeve using a lint-free cloth.
8.1.5.6

Wipe disinfection for the camera and mirror sleeve

NOTE
Do not spray the CEREC Omnicam or immerse it in cleaning agents
or disinfectants!
1. Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve  [→ 56]").
2. Use a new absorbent cotton gauze which is moistened with one of
the agents listed in the "Cleaning agents and disinfectants  [→ 52]"
section. First wipe the camera enclosure and then the mirror sleeve.
The camera enclosure and mirror sleeve are disinfected through
this step.
3. Next dry the mirror sleeve using a lint-free cloth.