Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect

Mirror sleeve for CEREC Omnicam
Pre-Cleaning process
(while mirror sleeve is attached to the camera)
1. Clean with a cotton gauze moistened with tap water of potable water quality that meets Federal Clean Water
Standards and pH neutral soap
or
use a clean lint-free cloth which has been soaked in CaviCide
or
use Cavi Wipes
or
use pdCARE Wipes
or
use a clean lint-free cloth which has been soaked in >60% isopropyl alcohol
2. Dry the mirror sleeve with a clean lint-free cloth.
HLD process
®
1. Use CIDEX
OPA as disinfectant for high-level
disinfection per manufacturer´s instructions.
2. Leave the mirror sleeve in CIDEX
12minutes at 20°C.
3. Rinse the sleeve under tap water of potable water
quality that meets Federal Clean Water Standards.
4. Dry the mirror sleeve with a clean lint-free cloth.
66 37 487 D3652
D3652.201.01.14.02 10.2017

Mirror sleeve reprocessing overview

8.1.3.5
See detailed instructions after summary.
®
OPA for
8.1.3.6

Pre-cleaning of the mirror sleeve

Cleaning process immediately after using the CEREC Omnicam: While
the mirror sleeve is attached to the camera, wipe off so that any surface
contamination cannot harden and adhere to the surface of the mirror
sleeve.
Use a cotton gauze moistened with tap water of potable water quality that
meets Federal Clean Water Standards and pH neutral soap or use a
clean lint-free cloth which has first been soaked in one of the cleaning
products listed in the „Cleaning and disinfection agents [ → 56]" section to
remove visible contamination.
Then dry the mirror sleeve with a clean lint-free cloth.
Dry heat sterilization process
➢ Dry heat 160°C (Sterident Model 200) for 120
minutes (wrapped or unwrapped)
or
Dry heat 160°C (SteriSURE) for 60minutes (wrapped
or unwrapped)
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 6
minutes (unwrapped),
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for
12minutes (wrapped).
8 Maintenance