Taking Acquisitions With The Cerec Omnicam - Dentsply Sirona CEREC AF Connect Operating Instructions Manual

Hide thumbs Also See for CEREC AF Connect:
Table of Contents

Advertisement

Available languages

Available languages

7 Operation
7.3 Taking acquisitions with the CEREC Omnicam
46
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
7.3

Taking acquisitions with the CEREC Omnicam

CAUTION
Hot surface!
The coated sapphire glass of the CEREC Omnicam is preheated in the
camera cradle. When removing the CEREC Omnicam from its cradle,
the surface temperature of the mirror sleeve can be up to 51°C. This
may cause an unpleasant heat sensation on contact with a person's skin
or mucous membrane. These temperatures will not damage the skin or
mucosal membrane.
After removing the CEREC Omnicam from the camera cradle, the
temperature of the mirror sleeve drops within a few minutes (< 5
minutes) to less than 43°C. The CEREC Omnicam is therefore suitable
for use in the patient's mouth for an unlimited period of time.
At an ambient temperature from 30°C, only select the three lower heater
settings.
The surface temperature of the mirror sleeve can reach up to 51°C
when the camera has been operating continuously for more than 20
minutes in an ambient temperature of more than 30°C.
NOTICE
Image brightness
The image brightness during the acquisition is controlled automatically,
so that there is always optimum image brightness, largely independent
of the distance between the CEREC AF/CEREC AI/CEREC AF
Connect/CEREC AI Connect and the tooth.
The surroundings of the tooth to be scanned should be as weakly
illuminated as possible. Avoid any type of external light. Switch off the
operating light.
IMPORTANT
Do not use cotton rolls in the scan area
Do not use any cotton rolls in the vicinity of the scan area.
Should any pieces of cotton roll contaminate this area, the acquisitions
will be inaccurate.
The teeth are blow-dried
1. Change to phase
 The camera is ready for scanning.
 A live image appears which can be used to look around the
patient's mouth.
"ACQUISITION"
.
D3652.201.01.14.02 10.2017
Dentsply Sirona
66 37 487 D3652

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerec aiCerec ai connect

Table of Contents