Set channel 02 to thermocouple type K and measure temperatures in the range –50.0 to
450.0°C. The measurable range for thermocouple type K is –200.0 to 1370.0°C.
The measured values in the range of –50.0 to 450.0°C are recorded in a width of 100
mm on the chart paper. This recording range is called a recording span , and the leftmost
and rightmost values of the recording span are called span left and span right ,
respectively.
−200.0°C
Measurable range of thermocouple type K
(Span left)
−50.0°C
100 mm
Entering Setting Mode
1.
Hold down the
Selecting the Range
2.
Press the
key with Range shown on the screen.
Set=Range
Input range and
Selecting the Channel Range
3.
Press the
key to set the first channel to 02 and then press the
First channel Last channel
CH=02-02
First channel 01-
4.
Likewise, set the last channel to 02 and then press the
Selecting the Input Type
5.
Press the
key to select TC and then press the
on page 37).
Mode=TC
Thermocouple
6.
Press the
key to select K and then press the
Range=K
R,S,B,K,E,J,T,N
Setting Span Left
7.
Set Span_L to –50.0 and press the
Press the
Press the
key to select the value.
Span_L= -50.0
-200.0/ 1370.0˚C
* When the
key or
reversed as when the respective key is pressed by itself.
450.0°C
Chart paper
key for 3 seconds to enter Setting mode.
MENU
Displays a description of the setup item.
Displays the thermocouple type.
key to select the desired digit.
key is pressed while holding down the SHIFT key, the operation is
1370.0°C
(Span right)
Displays the selectable range of channels.
key.
key (see "Explanation"
key.
key.
Displays the measurable range of
thermocouple type K.
key.