This document explains the firmware update procedure for the MT300 Digital Manometer (hereafter
referred to as the MT300).
When updating the firmware, be sure to follow this instruction manual.
Trademarks
• Microsoft, Windows 8.1, and Windows 10 are registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries.
• In this manual, the TM and ® symbols do not accompany their respective registered trademark or
trademark names.
• Other company and product names are trademarks or registered trademarks of their respective
holders.
Disclaimers
By downloading and installing this software, the customer agrees to all of the following disclaimers.
• Yokogawa bears no liability for any problems occurring as a result of downloading or installing this
software.
• Yokogawa bears no responsibility for any damage caused directly or indirectly as a result of using
this software.
• This software is provided free of charge, however no unlimited warranty against software defects
exists, nor is any claim made that the product is free of all defects whatsoever. Also, Yokogawa
is not always able to repair defects ("bugs") in, or respond to questions or inquiries about this
software.
• Yokogawa reserves all rights to this software, including but not limited to all property rights,
ownership rights, and intellectual property rights.

PC System Requirements

• Personal computer (PC)
A PC with a Core2 Duo 2 GHz or faster CPU and 1 GB or more of memory
• OS
Windows 8.1 or Windows 10

Notes before Updating

Setup Parameters
When you update the firmware, the MT300 setup parameters will be reset to their factory default
values. Save them if necessary.

Installing the USB Driver

You can update the firmware by connecting the MT300 to your PC with a USB cable and using the
firmware update tool.
To use the tool, YOKOGAWA's USB driver (YTUSB) must be installed in your PC. If it is not installed,
download it from the YOKOGAWA webpage.
For the installation procedure, see the manual that is downloaded with the driver.
Webpage:
File to download: USB driver (YTUSB)
MT300
Digital Manometer
Firmware Update Procedure
https://tmi.yokogawa.com/library/
1st Edition : May 2020 (YMI)
All Rights Reserved, Copyright © 2020 Yokogawa Test & Measurement Corporation
IM D027-01EN
1st Edition
1/3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MT300 and is the answer not in the manual?

Questions and answers

Summary of Contents for YOKOGAWA MT300

  • Page 1 Save them if necessary. Installing the USB Driver You can update the firmware by connecting the MT300 to your PC with a USB cable and using the firmware update tool. To use the tool, YOKOGAWA’s USB driver (YTUSB) must be installed in your PC. If it is not installed, download it from the YOKOGAWA webpage.
  • Page 2 It takes about 50 seconds for the update to complete. Do not remove the USB cable or turn off the MT300 until the update is finished. If you do, you may not be able to start the MT300. When updating completes successfully, the update tool closes automatically, and the MT300 restarts.
  • Page 3 Checking the MT300 Firmware Version When the pressure display screen appears, press MENU, and press the soft keys in the following sequence. Utility → Next 1/3 → Overview → Version Info The following product information screen appears. Check that the updated firmware version number is displayed next to Firmware Version. Firmware version IM D027-01EN...

Table of Contents