YOKOGAWA mR10000 436101 User Manual page 34

Hide thumbs Also See for mR10000 436101:
Table of Contents

Advertisement

Setting the Input Range and Alarm on Measurement Channels
34
Selecting the Input Type
5.
Press the
key to select 1-5V and then press the
on page 37).
Mode=1-5V
Scales and record
Setting Span Left
6.
Set Span_L to 1.000 and press the
Press the
Press the
key to select the value.
Span_L=
0.800/ 1.200V
Setting Span Right
7.
Likewise, set Span_R to 5.000 and press the
Span_R=
4.800/ 5.200V
Setting the Decimal Position and Scaling Left
8.
Display Scale_L.
Scale_L=
-200.00/ 300.00
9.
Press the
Scale_L=
-200.00/ 300.00
10.
Press the
key to select space and then press the
0.0).
Scale_L=
-200.00/ 300.00
Setting Scaling Right
11.
Likewise, set Scale_R to 500.0 and press the
Scale_R=
-2000.0/ 3000.0
The Setting complete screen is displayed. When this screen is displayed, the
settings entered up to then are applied.
Finishing the Settings
12.
When Setting complete screen is displayed, press the ESC key.
The Set=Range screen is displayed.
03-03 Channel
Setting complete
Setting the Unit
13.
Press the
key to select Unit and then press the
Set=Unit
Engineering unit
14.
Press the
key to set the first channel to 03 and then press the
First channel Last channel
CH=03-03
First channel 01-
15.
Likewise, set the last channel to 03 and then press the
* When the
key or
reversed as when the respective key is pressed by itself.
key.
key to select the desired digit.
1.000
Displays the range of Span_L.
5.000
Displays the range of Span_R.
0.00
Displays the scaling range.
key to select the desired digit.
0.00
0.0
500.0
Displays the scaling range.
Displays the description of setting item.
Displays the selectable range of channels.
key is pressed while holding down the SHIFT key, the operation is
key (see "Explanation"
key.
key (Scale_L is set to
key.
Set=Range
Input range and
key.
key.
key.
IM 04P01B01-02E