Commissioning

6.4 Installation Menus

4. Press <ENTER>. The current empty threshold number appears in a pop-up window.
5. Use the numeric keys to type a new Set Empty number.
6. To store the Set Empty number press <ENTER>
6.4
Installation Menus
FUH1010 Precision Volume Installation Menu Chart
Use <Left>, <Right>, <Up> and<Down> arrow buttons to navigate the menu between levels
and sub menus.
For example: To navigate to [Sensor Size]:
1. Press <MENU> to return to Level A.
2. Press <Right Arrow> to Level B (Dual Path Flow) then press <ENTER>.
3. Scroll using <Down Arrow> to [Install Sensor].
4. Press <Right Arrow> to Level D.
5. Scroll using <Down Arrow> to [Sensor Size].
6. Press <ENTER> select size from list.
Note
Menu items in bold are required entries to establish operation.
Level A
Level B
Meter Type
Dual Path
Flow
78
Level C
Level D (see
Chan/Path Setup
Recall Site Setup
Channel Enable
Create/Name Site Enter From List
Site Security
Delete Site
Save/Rename
Site
Pipe Data
Pick Pipe Class
Select Pipe Size
Pipe OD (in)
Pipe Material
Wall Thickness
Liner Material
Liner Thickness
ThermExpCoef
1/F
manual)
Enter From List
Enter From List
On/Off
Enter/Clear Site Name
Enter From List
Enter From List
Enter From List
Numeric Entry
Enter From List
Numeric Entry
Enter From List
Numeric Entry
Numeric Entry
Operating Instructions, 01/2013, A5E32122003-AA
Level E
FUH1010 IP65 NEMA 4X Precision Volume