Connecting
5.2 Transmitter Wiring
Power Supply
Power Supply Access Cover
Fuse F1
Input Power Conn. J10
Connector mounting screws
Figure 5-1
Input Power (J10) Wiring
4. Pull the desired length of input power wires through a cable gland and into transmitter
5. Wire input power connector for AC or DC power depending on power supply provided.
Note
Dress cables and make sure cable length is not excessive as to impede proper
replacement of access cover.
6. Insert wires into wire entry holes and secure by tightening wire clamp screws (see figure
Note
Power Supply connector wires should be stripped AWG 12 - 18 stranded wire or solid
conductors.
38
case.
above).
Wire Entry
Strip length 8mm (0.31 in)
Wire Clamp Screws
Ferrite power cord
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA