10. Disassemble the spacer bar and the unmounted frame. Use the bar as a straight edge
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
Sensor 1
Pipe
Dot
Sensor 2
and, with one edge against the mounted frames tapered roller center and the other
crossing the dot you drew, draw a line crossing the dot (see "B" above). Set the bar
aside.
Trim material from inner edge if necessary
Mylar Spacing Guide
8cm (3-inch) Overlapping Edge
Figure 5-14
Wrapping the Mylar Spacing Guide around the pipe (End View)
5.6 Sensor Installation
Sensor Edge Line
Spacer Bar
Dot
Line
Mylar Spacing Guide