Voltage Output Trim (Vo1 & Vo2)
Note
Can be trimmed to within .0025 V of nominal.
To calculate a voltage output:
1. Set up a multimeter to read volts, then connect it to the supply and return terminals of the
2. Move the highlight to the port to be tested by pressing the <Up/Down Arrow>. Then press
3. Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multimeter should now
4. If the multimeter reading does not match, use the numeric keys to type in the voltage
5. Press <ENTER> to register setting. This adjusts the flow meter's DAC (digital-to-analog
6. Re-check the multimeter to make sure that it is now reading 2.00 Volts.
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
voltage output under test.
the<Right Arrow> and then press <Down Arrow> to move the cursor to [Trim @ 2V].
be reading 2.00 Volts.
reading.
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multimeter.