Troubleshooting
10.3 Test Facilities Graph Screen
To activate the Test Facilities Graph Screen:
1. In the main menu, scroll to the [Diagnostic Data] menu and select [Test Facilities].
2. Scroll down to [Graph], press the <Right Arrow> and highlight [Yes]. Press <ENTER> to
3. The Test Facilities Graphic Screen will appear on the meter display as shown below.
Setting the Digital Damping Factor to a value HIGHER than the default value of 4 may be
necessary in cases where the signal-to-noise ratio (SN) is found to be unacceptably low
(<15:1), but only if the noise is determined to be asynchronous (i.e., not associated with the
transmit or flow meter timing circuitry) as shown in the signal example above, where the
baseline noise has a higher frequency than the true liquid signal.
The following application conditions may require a higher Digital Damping Factor:
● Close proximity to pressure control valves which may generate in-band acoustic noise
● High un-dissolved gas solids content in liquid.
● High electronic noise from variable frequency drives or other external equipment.
146
select.
Damping Factors
Digitized Receive Signal
Crossover Marker
Figure 10-2
Setting Digital Damping Factor
Min Damping Factor (Hot Key 1)
TN Marker
High Baseline Noise
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA