7. Observing the upstream and downstream orientation, attach the UP (upstream) and DN
8. Proceed to Commissioning (Page 73).
5.7
Mounting Temperature Sensors
Temperature is used to normalize the liquids sonic velocity in order to properly determine
interfaces and for density determination. Temperature sensors are available in clamp-on
style or in insert (Thermowell) style. Refer to the table below. Both styles incorporate 1000
ohm platinum RTD's for high precision.
Table 5- 2
Temperature Sensors
Description
Standard clamp-on RTD
Submersible clamp-on RTD
Insertion style RTD (size 1): 140mm (5.5 in)
Insertion style RTD (size 2): 216mm (8.5 in)
Insertion style RTD (size 3): 292mm (11.5 in)
Insertion style RTD (size 4): 368mm (14.5 in)
①
②
③
Figure 5-21
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
(downstream) cables to the sensors and make snug. Attach the other ends to the UP and
DN terminals of the transmitter.
7ME39600CR (992EC) Series Cable
Clamp-on Temperature Sensor
Thermal Couplant
Clamp-on Temperature Sensor
5.7 Mounting Temperature Sensors
Part Number
7ME39501TA00
7ME39501TB00
7ME39501TJ00
7ME39501TJ01
7ME39501TJ02
7ME39501TJ03
④
Pipe
⑤
Mounting Assembly
Connecting