7. Wait for EOT (End Of Transmission) to be displayed.
8. Close the file by pointing to Transfer, drag to Capture Text and click Stop button.
Closing the Terminal or HyperTerminal Program
You may now close the Terminal program. The file(s) you have downloaded are now saved
in the location you selected. You may now import the file you have saved into the
appropriate program (i.e. MS Word for site data, or MS Excel for Datalogger or wave shape
data for graphing or analysis).
The Datalogger contains data that has its fields separated by commas. By using the file
extension ".csv" (comma separated values) suggested earlier, the data will import directly
into MS Excel without any further modification. For the wave shape data, the fields are
separated by spaces, therefore, it is best to save those files as .txt and then use the MS
Excel Import Wizard to select "Space Delimiters" for importation of the data.
Site data is downloaded in plain text and can be imported directly into MS Word.
Reset Procedure using RS-232 port
SITRANS F 1010 systems allow you to perform a system reset via the RS-232 interface. The
following instructions require the meter to be connected serially to a PC.
Note
Custom RS-232 settings for baud rate, parity and data bits may not be preserved. Therefore,
be prepared to set your communications program back to the default (9600, Odd, 7) settings.
To Clear Active Memory using the RS-232 Interface
1. Turn off power (if it is currently on). Turn power on. As soon as you apply power,
2. Press the <Right Arrow> and then the <Down Arrow> to switch the option list to: [Clr
3. To restore operation, press <MENU> to access the Installation Menu. Create a new site
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
immediately type the @ character three times. The prompt: [Clr Active Memory? No]
appears at the top of the screen.
Active Memory? Yes] Press <ENTER> to clear all Active Site Data (but not saved site
setups).
setup or recall a stored site setup. Re-select any Meter Facilities items (e.g., RS-232
setup parameters).
Appendix A
A.3 RS-232 Connection
185