Troubleshooting
10.5 Force Transmit
Setting a Force Transmit condition
1. After [Install] command is invoked and while the flow meter is going through the drive
Note
The <ALT> and <MENU> keys must be pressed before the flow meter scans through all
the drives, or the Force Transmit function must be initiated again.
2. A typical menu screen will appear as shown below and indicate the current ALC (e.g.,
3. To exit Force Transmit, press the <Left Arrow> and a Detection Fault prompt will appear
4. Press the <Left Arrow> again and the meter will return to the [Install Sensor] menu and
158
selections press the <ALT> and <MENU> keys simultaneously.
50). This ALC number indicates the current receive signal strength and can be used for
further diagnostic purposes.
(see above).
highlight the [Empty Pipe Set] menu cell.
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA